Performance Tests
3- 50
RSA3408B Analyzer Service Manual
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408B
2.
Check spurious response:
a.
Modify the generator controls:
Set the generator output frequency to the value as shown in Table 3--7
(initially 25 MHz).
b.
Modify the analyzer controls:
H
Press the
Frequency/Channel
key on the front panel.
H
The
Center Freq
menu item is selected by default. Set the frequency
to the same value as the generator output using the numeric keypad.
c.
Measure the amplitude:
H
Press the
Marker Setup
key on the front panel.
H
Press the
Select Marker
side key to select
1
.
H
Press the
Peak
key and then the Markers:
B
(marker down) key
once to place the first marker at the peak on the noise floor.
H
Press the
Select Marker
side key to select
2
.
H
Press the
Peak
key to place Marker 2 on the signal peak.
H
Read the delta marker readout (
Δ
1--2) on the screen. Check that the
components other than the carrier meet the requirements shown in
Table 3--7 (initially --73 dBc).
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......