Removal and Installation Procedures
RSA3408B Analyzer Service Manual
5- 57
Figure 5- 28: Front panel assembly removal
1.
Assemble equipment and locate module to be removed:
a.
You will need a 12 mm nutdriver (Item 8), and a
1
@
16
inch hex wrench
(Item 12).
b.
Locate the rotary encoder in Figure 5--9 on page 5--19.
2.
Orient the front-panel assembly:
Set the front-panel assembly so its back is
down on the work surface and its front is facing you.
3.
Remove the general purpose knob:
See Figure 5--29. Loosen the setscrew
securing the knob using the
1
@
16
inch hex wrench. Pull the knob away.
4.
Remove the rotary encoder:
See Figure 5--29.
a.
Unplug the rotary encoder cable at J140 on the A60 Front key board.
b.
Using the 12 mm nutdriver, remove the nut securing the rotary encoder
to the front bezel.
5.
Reinstallation:
Reverse steps 3 and 4 to reinstall the rotary encoder and the
general purpose knob.
Rotary Encoder
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......