Adjustment Procedures
4- 34
RSA3408B Analyzer Service Manual
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408B
Figure 4- 11: Test hookup
c.
Initialize the analyzer:
H
Press the
System
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
d.
Modify the default settings:
H
Press the
Span
key on the front panel.
H
With the
Span
side key, set the span to 3 GHz using the numeric
keypad.
3.
Check the spectrum:
Confirm that a spectrum displays at 1.5 GHz.
4.
Disconnect the test equipment:
Disconnect the cable at
Input
.