Brief Procedures
RSA3408B Analyzer Service Manual
3- 5
Marker readout
Status indicator
Marker
Figure 3- 1: Spectrum of the calibration signal (100 MHz, - 20 dBm)
5.
Check the center frequency and peak amplitude using the marker:
a.
Press the
Peak
key on the front panel to place the marker on the peak
(see Figure 3--1).
b.
Check the marker readouts on screen. The frequency should be 100 MHz
and the amplitude should be approximately --20 dBm.
c.
Press
Marker Setup
>
Markers
>
Off
.
Check that the marker disappears.
6.
Check the RBW (Resolution Bandwidth) while changing the span setting.
a.
Press the
Span
key on the front panel.
b.
Confirm that the span is 36 MHz and the RBW is 100 kHz in the setup
display on the upper part of the screen (see Figure 3--2).
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......