Adjustment Procedures
RSA3408B Analyzer Service Manual
4- 29
Equipment
Required
One signal generator (Item 4)
Five 50
Ω
BNC coaxial cables (Item 9)
One 50
Ω
N-N coaxial cable (Item 10)
One GPIB cable (Item 11)
One RF attenuator (Item 13)
Prerequisites
Refer to page 4--1
Warm up the analyzer for more than one hour to stabilize the
bandpass filter (YTF).
1.
Prepare for the calibration:
See the following figure for the hookup.
GPIB cable
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408B
10 dB attenuator
10 MHz reference (BNC cable)
I+ (BNC cable)
I-- (BNC cable)
Q+ (BNC cable)
Q-- (BNC cable)
a.
Power off the RSA3408B analyzer and the signal generator.
b.
Connect the analyzer and the generator with GPIB cable.
c.
Connect the analyzer
Input
through a 10 dB attenuator followed by a 50
Ω
N-N coaxial cable to the generator output.
IF Flatness (Wide/IQ)
Calibration
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......