Performance Tests
3- 28
RSA3408B Analyzer Service Manual
Amplitude Checks
These procedures check the amplitude-related characteristics and are listed as
checked in
Specifications
of the
RSA3408B Technical Reference.
Equipment
Required
One signal generator (Item 1)
One power meter (Item 7)
One power sensor (Item 8)
One 50
Ω
N-N coaxial cables (Item 11)
Prerequisites
Refer to page 3--10.
1.
Set up the power meter and sensor:
NOTE
. Store the power sensor correction factors in the power meter, if you have
not yet done so.
a.
Connect the power meter and the power sensor. See the following figure.
Sensor input
Power meter
Power sensor
b.
Warm up the power meter and sensor for more than 20 minutes.
c.
Press
ZERO
on the power meter.
d.
Connect the RF input of the power sensor to the power reference output
of the power meter. See Figure 3--8 on page 3--29.
e.
Turn on
POWER REF
and execute the calibration.
f.
Disconnect the RF input of the power sensor from the reference output
of the power meter.
Check Absolute
Amplitude Accuracy
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......