Performance Tests
RSA3408B Analyzer Service Manual
3- 23
4.
Check frequency accuracy in the RF2 band:
a.
Modify the generator control:
Frequency
5 GHz
. . . . . . . . . . . . . . . .
b.
Modify the analyzer controls:
H
Press the
Frequency/Channel
key on the front panel.
H
The
Center Freq
menu item is selected by default.
Set the frequency to 5 GHz using the numeric keypad.
c.
Measure peak frequency:
H
Press the
Peak
key to place the marker on the signal peak.
H
Read the marker readout. Confirm that the value is
5 GHz
2.0 kHz.
5.
Check frequency accuracy in the RF3 band:
a.
Modify the generator control:
Frequency
7 GHz
. . . . . . . . . . . . . . . .
b.
Modify the analyzer controls:
H
Press the
Frequency/Channel
key on the front panel.
H
The
Center Freq
menu item is selected by default.
Set the frequency to 7 GHz using the numeric keypad.
c.
Measure peak frequency:
H
Press the
Peak
key to place the marker on the signal peak.
H
Read the marker readout. Confirm that the value is
7 GHz
2.4 kHz.
6.
Disconnect the test equipment:
Disconnect the cable at the analyzer input.
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......