Performance Tests
RSA3408B Analyzer Service Manual
3- 43
Equipment
Required
None
Prerequisites
Refer to page 3--10.
1.
Prepare for the test:
a.
Initialize the analyzer:
H
Press the
System
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
RSA3408B
b.
Modify the analyzer controls:
H
Press the
Frequency/Channel
key on the front panel.
H
The
Center Freq
menu item is selected by default.
Set the frequency to 10 MHz using the general purpose knob.
H
Press the
Span
key on the front panel.
H
The
Span
menu item is selected by default.
Set the span to 1 MHz using the general purpose knob.
H
Press the
Amplitude
key on the front panel.
H
The
Ref Level
menu item is selected by default.
Set the level to --30 dBm using the general purpose knob.
H
Press the
RBW/FFT
key on the front panel.
H
Press the
RBW/FFT
side key to select
Man
.
H
The
RBW
menu item is selected by default.
Set the value to 10 kHz using the general purpose knob.
Check Displayed Average
Noise Level
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......