![Tektronix RSA3408B Service Manual Download Page 122](http://html1.mh-extra.com/html/tektronix/rsa3408b/rsa3408b_service-manual_1078148122.webp)
Adjustment Procedures
4- 32
RSA3408B Analyzer Service Manual
d.
Set the GPIB address of the generator:
H
Press the
Config...
side key.
H
Press the
AWG Address
side key and set the generator address.
H
Press the
Cancel--Back
side key.
3.
Perform the IQ gain calibration in all bands:
a.
Press the
Target Files
side key to select
All
.
b.
Press the
IQ Gain
side key to perform the IQ gain calibration.
NOTE
. You can select the calibration items and bands as described in
Service
Cal Menu
on page 4--20.
4.
Disconnect the test equipment:
a.
Disconnect the cable from the analyzer input.
b.
Turn off the analyzer and generators.
c.
Disconnect the GPIB cables from all the ports.
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......