Performance Tests
RSA3408B Analyzer Service Manual
3- 33
Equipment
Required
One signal generator (Item 1)
One 50
Ω
N-N coaxial cables (Item 11)
Prerequisites
Refer to page 3--10.
1.
Prepare for the test:
a.
Set the generator controls:
Frequency
100 MHz
. . . . . . . . . . . . . . . .
Amplitude
--25 dBm
. . . . . . . . . . . . . . . .
RF output
On
. . . . . . . . . . . . . . . .
b.
Hook up the generator:
Connect the generator output through a 50
Ω
N-N coaxial cable to the
analyzer
Input
. See the following figure.
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408B
c.
Initialize the analyzer:
H
Press the
System
key on the front panel.
H
Press the
Reset All to Factory Defaults
side key.
Check Input Attenuator
Setting Uncertainty
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......