Performance Tests
RSA3408B Analyzer Service Manual
3- 37
3.
Determine the reference power in 0 to --30 dBm:
a.
Measure the power and record it as Pr0.
b.
Set the signal generator amplitude from 0 to --30 dBm in 10 dB steps and
measure the power as Pr1 to 4, respectively.
Signal generator amplitude Reference power
+10 dBm
Pr0
0 dBm
Pr1
--10 dBm
Pr2
--20 dBm
Pr3
--30 dBm
Pr4
c.
Disconnect the power sensor input from the generator.
4.
Modify the hookup:
a.
Connect the generator output through a 50
Ω
N-N coaxial cable to the
spectrum analyzer (Item 5) input. See the following figure.
50
Ω
N-N coaxial cable
Output
Signal generator
Spectrum analyzer
b.
Set the signal generator control:
Amplitude
--30 dBm
. . . . . . . . . . . . . . . .
c.
Set the spectrum analyzer controls
:
Measurement mode
SA mode
. . . . . . . .
Reference level
--20 dBm
. . . . . . . . . . . .
Center frequency
100 MHz
. . . . . . . . . .
Span
1 MHz
. . . . . . . . . . . . . . . . . . . .
RF attenuation
10 dB
. . . . . . . . . . . .
RBW
Auto
. . . . . . . . . . . . . . . . . . . .
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......