Adjustment Procedures
RSA3408B Analyzer Service Manual
4- 5
2.
Install the hookup and preset the instrument controls:
a.
Turn off the power meter and the analyzer.
b.
Connect the GPIB cables on the power meter and the analyzer rear panel.
c.
Connect a mouse and a keyboard to the analyzer.
d.
Turn on the power meter and the analyzer.
e.
Set the GPIB address of the power meter to 13 (fixed).
GPIB cable
Power meter
RSA3408B
Figure 4- 1: Hookup for storing the cal factors in the power meter
3.
Store the calibration factors:
a.
If you use a floppy disk, insert it to the disk drive on the analyzer.
b.
Press
Cal
>
Service...
>
Password
.
c.
Enter
270833
and then press
Enter
using the numeric keypad.
d.
Press the
Cal
side key.
e.
Press
Measure
>
SG Flatness
.
f.
Press the
Read Cal Factor
side key.
The file selection dialog box appears.
g.
Enter the file name and then press the
Open
button.
The calibration factors are stored in the power meter.
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......