Performance Tests
3- 40
RSA3408B Analyzer Service Manual
Spurious Response Checks
These procedures check the spurious-related characteristics and are listed as
checked in
Specifications
of the
RSA3408B Technical Reference.
Equipment
Required
Two signal generators (Item 2 (or 3) and Item 4)
One power combiner (Item 9)
Three 50
Ω
N-SMA coaxial cables (Item 12)
Prerequisites
Refer to page 3--10.
1.
Prepare for the test:
a.
Connect each generator output through a 50
Ω
N-SMA coaxial cable to a
power combiner input. See the following figure.
Signal generator
RSA3408B
Output
Signal generator
Power
combiner
N-SMA cable
N-SMA cable
Output
b.
Connect the power combiner output through a 50
Ω
N-SMA coaxial
cable to the analyzer
Input
.
Check 3
rd
Order
Intermodulation Distortion
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......