Performance Tests
3- 38
RSA3408B Analyzer Service Manual
5.
Set the reference offset:
a.
Set the spectrum analyzer control:
Average
10
. . . . . . . . . . . . . . . . . .
b.
Measure the peak power Ppr using the maker peak.
c.
Modify the spectrum analyzer control:
Reference offset
Pr4 -- Ppr
. . . . . . . . . . .
6.
Determine the reference power in --40 to --50 dBm:
Set the signal generator amplitude to --40 and --50 dBm to measure the power
as Pr5 and 6, respectively.
Signal generator amplitude Reference power
--40 dBm
Pr5
--50 dBm
Pr6
a.
Disconnect the cable from the spectrum analyzer.
7.
Modify the hookup:
a.
Connect the signal generator output to the RSA3408B analyzer input.
See the following figure.
50
Ω
N-N coaxial cable
Output
Signal generator
RSA3408B
b.
Set the signal generator control:
Amplitude
+10 dBm
. . . . . . . . . . . . . . . .
Summary of Contents for RSA3408B
Page 4: ......
Page 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Page 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Page 19: ...Specifications...
Page 20: ......
Page 22: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Page 23: ...Theory of Operation...
Page 24: ......
Page 30: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Page 31: ...Performance Verification...
Page 32: ......
Page 88: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Page 89: ...Adjustment Procedures...
Page 90: ......
Page 127: ...Maintenance...
Page 128: ......
Page 209: ...Diagrams...
Page 210: ......
Page 212: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Page 214: ...6 4 RSA3408B Analyzer Service Manual...
Page 216: ...6 6 RSA3408B Analyzer Service Manual...
Page 217: ...Replaceable Parts List...
Page 218: ......
Page 249: ......
Page 250: ......