Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-74
077114601 / March 2019
:BATTery[n]:TRACe:TRIGger:VOLTage:STATe
This command sets and queries the state of voltage level trigger.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0 (OFF)
Usage
:BATTery[n]:TRACe:TRIGger:VOLTage:STATe <state>
:BATTery[n]:TRACe:TRIGger:VOLTage:STATe?
[n]
Channel number: For single-channel instruments, the value is 1
<state>
ON or 1: Enable the voltage level trigger
OFF or 0: Disable the voltage level trigger
Example
:BATT:TRAC:TRIG:VOLT:STAT 1
Enable the voltage level trigger.
Also see
:BATTery[n]:TRACe:TRIGger:VOLTage[:LEVel]
(on page 7-73)
:BATTery[n]:TRACe:TRIGger:VOLTage:DIRection
(on page 7-73)
:BATTery[n]:OUTPut:PROTection:CLEar
This command clears the overvoltage, overcurrent, overtemperature, or reverse fault status.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:OUTPut:PROTection:CLEar
[n]
Channel number: For single-channel instruments, the value is 1
Details
This command clears the protection status if an overvoltage, overcurrent, or overtemperature
protection condition occurred or a reverse sense fault was detected.
For all four faults, this command also clears the related Questionable Instrument Summary Condition
Register bit.
Example
:BATT:OUTP:PROT:CLE
Clear the fault status.
Also see
:BATTery[n]:OUTPut:PROTection:TRIPped?
(on page 7-75)