Section 2: General operation
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
2-100
077114601 / March 2019
To select measurement autoranging using the front-panel Measure Settings screen:
1. Press the
MENU
key.
2. Under Measure, select
Settings
.
3. Select the button next to
Range
. The Range dialog box is displayed.
4. Select
AUTO
.
You can set specific measurement ranges or allow the instrument to choose the ranges
automatically (power supply function only).
To set the current measurement range, send the command:
:SENSe:CURRent:RANGe:AUTO ON
To set the concurrent measurement range, send the command:
:SENSe:CONCurrent:RANGe:AUTO ON
You can set specific measurement ranges or allow the instrument to choose the ranges
automatically (power supply function only).
Protection
The 2281S provides several methods for ensuring that the source remains within certain values. This
helps to protect the device under test (DUT) from damage.
The protection settings that affect the source are:
(on page 2-101). This is the voltage at the instrument terminals.
(on page 2-102). This is the current at the instrument terminals.
(on page 2-104). This is the sourced voltage value at the device.
The 2281S also provides overtemperature protection to prevent damaging heat build-up.