This subsystem contains commands that affect the overall operation of the instrument, such as
autozero, beeper, communications, event logs, and time.
:SYSTem:AZERo[n][:STATe]
This command enables and disables the autozero.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1 (ON)
Usage
:SYSTem:AZERo[n][:STATe] <state>
:SYSTem:AZERo[n][:STATe]?
[n]
Channel number: For single-channel instruments, the value is
1
<state>
Disable autozero:
OFF
or
0
Enable autozero:
ON
or
1
Details
When autozero is disabled, the reference and internal temperature measurements are not performed.
This increases the measurement speed. However, the reference and temperature reference points
will eventually drift, resulting in inaccurate readings of the input signal. For details, refer to
(on page 4-5).
Example
:SYST:AZER OFF
Disables autozero to increase measurement speed.
Also see