Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-51
:BATTery[n]:TEST:MEASure:EVOC?
This command queries the values of ESR and Voc when the power supply is in Battery Test mode.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:SIMulator:MEASure:EVOC?
[n]
Channel number: For single-channel instruments, the value is 1
Details
When this command is executed, the instrument starts a new measurement and returns the values of
ESR and Voc. Output is off after the measurement.
Example
:BATT:TEST:MEAS:EVOC?
The values of ESR and Voc are returned.
Also see
:BATTery[n]:TEST:SENSe:EVOC:DELAy
(on page 7-56)
:BATTery[n]:TEST:SENSe:AH:ESRInterval
This command sets and queries the ESR sample interval during A-H measurement in the battery test function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
S10
Usage
:BATTery[n]:TEST:SENSe:AH:ESRInterval <interval>
:BATTery[n]:TEST:SENSe:AH:ESRInterval?
[n]
Channel number: For single-channel instruments, the value is 1
<interval>
S10 = 10 seconds
S30 = 30 seconds
S60 = 60 seconds
S120 = 120 seconds
M10 = 10 minutes
Example
:BATT:TEST:SENS:AH:ESRI S60
:BATT:TEST:SENS:AH:ESRI?
Set the ESR sample interval to 60 s and query the
value.
Return: S60
Also see
None