Section 5: Application examples
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
5-4
077114601 / March 2019
Set up remote communications
You can run this application from the front panel or any of the supported communication interfaces for
the instrument (GPIB, USB, or ethernet).
The following figure shows the rear-panel connection locations for the remote communication
interfaces. For additional information about setting up remote communications, see
(on page 2-68).
Figure 116: 2281S remote interface connections
Device connections
Connect a 1 kΩ resistor to the output binding posts on the front panel. For details, refer to
(on page 2-63). You also can use four-wire connections to compensate for
voltage drops on the load leads.
Configure a 10-step linear list sweep
To configure a list on the front panel:
1. In the startup screen, choose the power supply function.
2. Press the
MENU
key.
3. Under Measure, highlight the
List
icon using either the navigation control or the soft keys. Press
ENTER
.
4. Select point 1 and press
ENTER
to select the box.