Section 5: Application examples
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
5-6
077114601 / March 2019
Execute a 10-step linear list sweep
To execute a list sweep on the front panel:
1. In the startup screen, choose the power supply function.
2. Press the
MENU
key.
3. Under Trigger, highlight the
Configure
icon using either the navigation control or the soft keys.
Press
ENTER
.
4. Select the button next to
Sample Count
and press the
ENTER
key.
5. Set the sample count to 10.
6. Press the
HOME
key.
7. Press the
NEXT
soft key twice to navigate to the third home screen.
8. Press the
List
soft key. The List dialog box is displayed.
9. If the
Number
soft key is not showing a 1, press the
Number
soft key and enter the value, 1.
10. Press the
Enable
soft key to run list 1.
To view the measurements on the front panel:
1. Press the
Exit
key to close the List dialog box.
2. Press the
View
soft key under Graph to view a plot of the measurements.
Figure 118: Graph view of 10-step linear list sweep