Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-67
:BATTery[n]:TRACe:TRIGger:CURRent:STATe
This command sets and queries the state of current level trigger.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0 (OFF)
Usage
:BATTery[n]:TRACe:TRIGger:CURRent:STATe <state>
:BATTery[n]:TRACe:TRIGger:CURRent:STATe?
[n]
Channel number: For single-channel instruments, the value is 1
<state>
ON or 1: Enable the current level trigger
OFF or 0: Disable the current level trigger
Example
:BATT:TRAC:TRIG:CURR:STAT 1
Enable the current level trigger.
Also see
:BATTery[n]:TRACe:TRIGger:CURRent[:LEVel]
(on page 7-66)
:BATTery[n]:TRACe:TRIGger:CURRent:DIRection
(on page 7-66)
:BATTery[n]:TRACe:CLEar
This command clears all the data in the buffer.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:TRACe:CLEar
[n]
Channel number: For single-channel instruments, the value is 1
Example
:BATT:TRAC:CLE
Clear the data buffer.
Also see
None