Section 3: Functions and features
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
3-52
077114601 / March 2019
Sweep mode triggering
In sweep mode, the instrument generates a measure complete signal after each list cycle. You can
detect this signal in digital I/O line 2. For details, refer to
(on page 3-22).
Figure 93: Sweep mode triggering
As shown in the previous figure, the device action can be any mode described in
page 3-48). The measurement time is set accordingly.