Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-93
:CALibration:PROTected:DATA?
This command queries the saved calibration constants.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Nonvolatile memory
Not applicable
Usage
:CALibration:PROTected:DATA?
Details
(on page C-31).
Example
:CAL:PROT:DATA?
Return the calibration constants in order:
+4.989000E-02,+4.00,+9.091379E
-01,+6.047600E-01,+1.306331E-02,+5.688
220E+01,+1.00,+1.521200E-03,+4
.262407E-03,+4.687520E-03,+1.500000E-0
1,+3.965609E-01,+3.980984E-03,+3.51767
5E-01,+5.200000E-03,+1.376206E-02,+1.4
58743E-02,+5.031000E-01,+1.00
Also see
None
:CALibration:PROTected:DATE
This command specifies the calibration date.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
Not applicable
Usage
:CALibration:PROTected:DATE <year>,<month>,<day>
:CALibration:PROTected:DATE?
<year>
Year; must be more than 1970
<month>
Month: 1 to 12
<day>
Day: 1 to 31
Details
The query command returns the latest calibration date in the following format:
<year> <month>
<day>