Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-86
077114601 / March 2019
:CALCulate2:FORMat:ELEMents
This command specifies or queries data elements to be included in the data string. It only influences minimum
and maximum query results.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
READ, UNIT
Usage
:CALCulate2:FORMat:ELEMents <itemList>
:CALCulate2:FORMat:ELEMents <itemList>,<channel>
:CALCulate2:FORMat:ELEMents?
:CALCulate2:FORMat:ELEMents? <channel>
<itemList>
The available data elements:
Voltage and current reading:
READing
Reading units:
UNIT
Reading number:
RNUMber
Timestamp:
TSTamp
. For example, output might be:
07/01/2013
19:21:36.2553
<channel>
The channel number (default value is CH1):
Channel 1: CH1
Channel 2: CH2
Details
This command affects the output of the following commands:
:CALCulate2:IMMediate?
:CALCulate2:DATA?
You can specify one or more elements.
<itemList>
is a string that must be enclosed in single or
double quotes. Each element in the list must be separated by a comma (,).