Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-173
:SYSTem:PASSword:NEW
This command sets a new password for the remote connection.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
admin
Usage
:SYSTem:PASSword:NEW <string1>,<string2>
<string1>
A string that contains the present password (maximum 8 characters)
<string2>
A string that contains a new password (maximum 8 characters)
Details
When the access to the instrument is set to lock, this is the password that is used to gain access.
The instrument continues to use the old password for all interactions until the command to change it
executes. When changing the password, give the instrument time to execute the command before
attempting to use the new password.
If you forget the password, you can reset the password to the default. On the front panel, press
MENU
. Under System, select Manage. Select
Password Reset.
In the pop-up window, select
LXI
to
reset the password.
Example
:SYST:PASS:NEW "admin","123456"
Change the password of the instrument to 123456
Also see
(on page 7-172)
(on page 7-174)