:BATTery[n]:TRACe:TRIGger:OCCur?
This command queries whether a level trigger has occurred both in the battery simulator and battery test function.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:BATTery[n]:TRACe:TRIGger:OCCur?
[n]
Channel number: For single-channel instruments, the value is 1
Example
:BATT:TRAC:TRIG:OCC?
Queries whether a level trigger has occurred.
If a level trigger has occurred, the value of 1 will be
returned.
Also see
None
:BATTery[n]:TRACe:TRIGger:OFFSet
This command defines the number of points at which the instrument begins to log data in the buffer both in the
battery simulator and battery test function.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:BATTery[n]:TRACe:TRIGger:OFFSet <offset>
:BATTery[n]:TRACe:TRIGger:OFFSet?
[n]
Channel number: For single-channel instruments, the value is 1
<offset>
0 to 2500
Example
:BATT:TRAC:TRIG:OFFS 10
:BATT:TRAC:TRIG:OFFS?
10 points before the level trigger activation are cleared.
Also see
None