Section 3: Functions and features
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
3-28
077114601 / March 2019
For the battery test function
Digital I/O line
Function
Action
1
Trigger in, manual output high, manual
output low, and manual input
—
2
Digital out, digital in, manual output high,
manual output low, and manual input
Digital Out:
Sink<>Source, Hi V, Low V,
CC<>CV, End
Digital In:
Output On, Output Off, Script
Start, Script Stop
3
Fault output, manual output high, manual
output low, and manual input
—
4
Digital out, digital in, manual output high,
manual output low, and manual input
Digital Out:
Sink<>Source, Hi V, Low V,
CC<>CV, End
Digital In:
Output On, Output Off, Script
Start, Script Stop
5
Digital out, digital in, manual output high,
manual output low, and manual input
Digital Out:
Sink<>Source, Hi V, Low V,
CC<>CV, End
Digital In:
Output On, Output Off, Script
Start, Script Stop
6
Digital out, digital in, manual output high,
manual output low, and manual input
Digital Out:
Sink<>Source, Hi V, Low V,
CC<>CV, End
Digital In:
Output On, Output Off, Script
Start, Script Stop
For the battery simulator function
Digital I/O line
Function
Action
1
Trigger in, manual output high, manual
output low, and manual input
—
2
Digital out, digital in, manual output high,
manual output low, and manual input
Digital Out:
Sink<>Source, Hi V, Low V,
Full, Empty
Digital In:
Output On, Output Off
3
Fault output, manual output high, manual
output low, and manual input
—
4
Digital out, digital in, manual output high,
manual output low, and manual input
Digital Out:
Sink<>Source, Hi V, Low V,
Full, Empty
Digital In:
Output On, Output Off
5
Digital out, digital in, manual output high,
manual output low, and manual input
Digital Out:
Sink<>Source, Hi V, Low V,
Full, Empty
Digital In:
Output On, Output Off
6
Digital out, digital in, manual output high,
manual output low, and manual input
Digital Out:
Sink<>Source, Hi V, Low V,
Full, Empty
Digital In:
Output On, Output Off
When you set the arm source or trigger source to external, digital I/O line 1 is set to trigger in
automatically.