:TRIGger[:SEQuence[n]]:SAMPle:COUNt
This command sets the number of measurement per trigger event.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
1
Usage
:TRIGger[:SEQuence[n]]:SAMPle:COUNt <count>
:TRIGger[:SEQuence[n]]:SAMPle:COUNt?
:TRIGger[:SEQuence[n]]:SAMPle:COUNt? DEFault
:TRIGger[:SEQuence[n]]:SAMPle:COUNt? MINimum
:TRIGger[:SEQuence[n]]:SAMPle:COUNt? MAXimum
[n]
Channel number: For single-channel instruments, the value is
1
<count>
The number of measurements: 1 to 2500
Details
This command sets the number of measurement per trigger event. For details, refer to
(on page 3-30).
Example
:TRIG:SAMP:COUN 10
Set the sample count to 10.
Also see
(on page 3-30)