Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 3: Functions and features
077114601 / March 2019
3-41
7. After adding points to the list, select the Voltage box, Current box or Dwell box in the specified
point row. Press the
ENTER
key. The editing window is displayed. You can use the numerical
keys or the navigation control to change the value.
8. Select the
Save
button to store the list in memory location.
9. Enter a name for the list on the displayed keyboard and press
ENTER
.
To configure a list using SCPI commands:
To set up list 1 with 6 points listed in the following table, send these commands:
:LIST1:CURR "1,1,1"
:LIST1:VOLT "5,10,15"
:LIST1:DWEL "2,2,2"
:LIST1:CURR:APPE "2,2,2"
:LIST1:VOLT:APPE "5,10,15"
:LIST1:DWEL:APPE "3,3,3"
:LIST:SAVE:INTE 1
Points
Voltage (V)
Current (A)
Dwell (s)
1
5
1
2
2
10
1
2
3
15
1
2
4
5
2
3
5
10
2
3
6
15
2
3
Inserting or copying points from the front panel
To insert or copy one or more points from the front panel:
1. Highlight the Points cell of the row you want to insert.
2. Press
ENTER
to select the cell row. You can repeat steps 1 and 2 to select multiple points.
3. Select
Insert
or
Copy
and press
ENTER
.
Deleting points from the front panel
To delete one or more points from the front panel:
1. Highlight the Points cell of the row you want to delete.
2. Press
ENTER
to select the cell row. You can repeat steps 1 and 2 to select multiple points.
3. Select
Delete
and press
ENTER
.