Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Appendix C: Calibration and adjustment
077114601 / March 2019
C-19
Step 12
1. Before starting this calibration step, configure the shunt and load (electronic load or resistor load)
as shown in the following table.
Model
Shunt
Power supply
2281S
5.05 Ω
Constant current mode
5.0 V
1.0 A
2. Enter the measured result from the digital multimeter (DMM) and press the
ENTER
key. After
entering the data, the waiting window is displayed. You need to wait a few seconds for the
internal measurements and calculations. The
Save
calibration window is displayed.
3. Press
Save
to save all the calibration constants to system internal memory. If you press
Cancel
,
you will lose all the calibration constants and exit the calibration.
Saving calibration constants from the front panel
1. Press the
Save
soft key. The Enter Date window is displayed.
2. Enter the present date and press the
ENTER
key. The date and calibration constants are saved
to instrument's internal memory.