54
Table 202
G
uidance and manufacturer
’s
declaration – electromagnetic immunity
UNIDRIVE
®
G
YN model
20
7111 20-1 i
s
intended for u
s
e in the electromagnetic environment
s
pecified below. The cu
s
tomer or u
s
er of the UNIDRIVE
®
G
YN model
20
7111 20-1
s
hould en
s
ure that it i
s
u
s
ed in
s
uch an environment.
Immunity te
s
t
EN/IEC 60601 te
s
t level
Compliance level
Electromagnetic environment - guidance
Electro
s
tatic Di
s
charge (E
S
D)
IEC 61000-4-2
± 6
k
V contact
± 8
k
V air
Complie
s
± 6
k
V contact
± 8
k
V air
Floor
s
s
hould be wood, concrete or ceramic tile. If
floor
s
are covered with
s
ynthetic material, the relative
humidity
s
hould be at lea
s
t 30%.
Electrical fa
s
t
tran
s
ient/bur
s
t
IEC 61000-4-4
± 2
k
V for power
s
upply line
s
± 1
k
V for input/output line
s
Complie
s
± 2
k
V for power
s
upply
line
s
± 1
k
V for input/output line
s
Main
s
power
q
uality
s
hould be that of a typical
commercial or ho
s
pital environment. If image
di
s
tortion occur
s
it may be nece
ss
ary to further filter
the power
q
uality by u
s
ing a AC line filter.
S
urge
IEC 61000-4-5
± 1
k
V differential mode
± 2
k
V common mode
Complie
s
± 1
k
V differential mode
± 2
k
V common mode
Main
s
power
q
uality
s
hould be that of a typical
commercial or ho
s
pital environment.
Voltage dip
s
,
s
hort
interruption
s
and voltage
variation
s
on power
s
upply
input line
s
IEC 61000-4-11
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycle
s
70% U
T
(30% dip in U
T
)
for 25 cycle
s
<5% U
T
(>95% dip in U
T
)
for 5
s
ec
Complie
s
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycle
Complie
s
40% U
T
(60% dip in U
T
)
for 5 cycle
s
Complie
s
70% U
T
(30% dip in U
T
)
for 25 cycle
s
Complie
s
<5% U
T
(>95% dip in U
T
)
for 5
s
ec
Main
s
power
q
uality
s
hould be that of a typical
commercial or ho
s
pital environment.
If the u
s
er of the e
q
uipment or
s
y
s
tem re
q
uire
s
continued operation during power main
s
interruption
s
,
it i
s
recommended that the e
q
uipment or
s
y
s
tem be
powered from an uninterruptible power
s
upply or a
battery.
Power fre
q
uency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Complie
s
3 A/m
Power fre
q
uency magnetic field
s
s
hould be at level
s
characteri
s
tic of a typical location in a typical
commercial or ho
s
pital environment.
* Note:
U
T
i
s
the a.c. main
s
voltage prior to application of the te
s
t level.
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Ïðèëîæåíèå
Óêàçàíèÿ ïî ýëåêòðîìàãíèòíîé
ñîâìåñòèìîñòè (ÝÌÑ)
Appendix
Electromagnetic Compatibility (EMC)
Information