Page 2-4
IsoAir 310P Aerosol Particle Sensor Operations Manual
Chapter 2 Unpacking and Installation
Sensor Installation
Setting Up a Remote Sampling System
It is not necessary to position the IsoAir 310P enclosure in the air flow to be sampled.
If needed, the enclosure can be installed in a remote location. Connect up to 3 meters
of sample tubing from the enclosure to an isokinetic probe (ISP), positioning the ISP in
the airflow sampling area.
NOTE:
In pharmaceutical applications, no more than 2 meters of inlet
tubing is recommended.
Required materials:
• Bev-a-Line 3/8-inch ID sampling tube
3 meters long (Provided)
• IsoAir 310P ISP (Provided)
• A standard laboratory mounting system for ISP (User must provide)
>> To set up an IsoAir 310P for remote sampling:
1
Set up a mounting system in the area to be monitored.
2
Remove the ISP from the IsoAir 310P enclosure and connect it to the mounting
system.
3
Insert one end of the Bev-a-Line tubing through the hole in the top of the IsoAir 310P
enclosure and over the particle counter inlet tube.
4
Attach the other end of the tube to the bottom of the ISP.
5
Adjust the ISP on the laboratory mounting system to achieve the best orientation for
sampling. ISP should be oriented in accordance with the air flow direction.
Installing the Sensor
The following instructions are provided as a guide to help you install the IsoAir 310P
particle sensor:
• Removing the Electronics Module on page
• Drilling Holes for Conduit Fittings (as needed) on page
2-9(as needed)
• Connecting the Electrical Wiring on page