Configuring the Software
Chapter 4 Facility Net Setup
IsoAir 310P Aerosol Particle Sensor Operations Manual
Page 4-3
Figure 4-2 IsoAir
310P Configuration
dialog box
5
Complete the following fields in the
IsoAir
310P Configuration
dialog box:
• Units
• Minimum Interval
• Maximum Interval
• Name
Table 4-1 Configuration dialog box field definitions
Units
Use the pull-down menu to select the units to display. Cubic feet is “cf”
and is the default unit presented by the IsoAir 310P.
Minimum Interval
Defines the time in seconds that the IsoAir 310P particle sensor will
process an individual sample. If the maximum sample interval is the
same as the minimum sample interval, all samples will be processed as
reported by the instrument. The volume sampled over time depends on
the flow rate. The sensor uses this interval when particles are being
counted. When no particles are counted, the Maximum Interval is used.
This allows for reduced data storage space.
Maximum Interval
If the maximum sample interval is greater than the minimum sample
interval, zero count reports from individual sensors will be
accumulated (in time) up to the maximum interval. The maximum
sample interval is always a multiple of the minimum sample interval.
The volume sampled over time depends on the flow rate. This sample
interval is used by the IsoAir 310P particle sensor when zero particles
are being counted, to reduce data storage space requirements.
Name
Enter a name for the sample point here. The name must be unique to
the system. Facility Net will notify you if a chosen name is already in
use.