Page 2-12
IsoAir 310P Aerosol Particle Sensor Operations Manual
Chapter 2 Unpacking and Installation
Sensor Installation
Connecting the Electrical Wiring
CAUTION
• All wiring must be done in accordance with applicable electrical codes.
• A qualified electrician should be used for wiring AC Mains.
• This instrument requires an external power disconnect for safety.
Electrical wiring should be rated to handle the following current, per instrument:
• 100 – 240 V, 50 – 60 Hz, 1 A
Wires connecting to the AC terminal block must be 14 AWG / 2.5 mm
2
minimum. See
for wiring placement. The terminal block can be removed by loosening its
two screws and rotating/sliding out of slots.
The external safety disconnect may be either a switch or a circuit breaker. A single or
double pole switch is acceptable.
Figure 2-9
AC Input Wiring Location on Rear
>> To install the wiring:
1
Install wiring through the conduit port(s).
2
Connectors must be field-wired to the terminal block as shown in
3
Terminate the ground wire to the green ground screw near the terminal block.