OPC Setup
Appendix E OPC Configuration and Operation
IsoAir 310P Aerosol Particle Sensor Operations Manual
Page E-3
Per the OPC Data Access server specification, each presented data tag will have an
associated time stamp. Since the IsoAir 310P particle sensor has no real-time clock
and no way to synchronize time, the time stamp will be accurate in its chronological
progression, but the absolute time/date indicated will be arbitrary.
Quality on the counter tags is set bad if flow and/or laser status is bad. All other tags
will have quality set good since the data presented by these tags are still valid.
After powering on, the particle sensor is constantly sampling. Data is made available
at the end of each defined sample interval.
The “
Set
Resynch
” command is set by selecting 0 or 1. This allows the counter to
synchronize its sample with the first OPC connection.
OPC Setup
IsoAir 310P OPC provides data via an OPC Server. The server supports OPC Data Access
Custom Interface Standard 2.05a. Information on the OPC specifications and
standards can be found at http://www.opcfoundation.org/.These specifications and
standards are essential for understanding how to use OPC.
Sample.FlowRate
VT_R4
cf/min
R
Sample flow rate
Sample.Interval
VT_R4
Seconds
R
Sample interval
Sample.Volume
VT_R4
cf
R
Sample volume
Sizes.Ch1
VT_R4
Microns
R
Channel sizes - fixed
Sizes.Ch2
VT_R4
Microns
R
Sizes.Ch3
VT_R4
Microns
R
Sizes.Ch4
VT_R4
Microns
R
Status.State
VT_BSTR
None
R
System status: Example is "Idle" or
"Sampling"
Status.String
VT_BSTR
None
R
Sample status: Example is "Good Laser &
Good Flow"
Status.Value
VT_I2
Status
R
Sample status: Encoded flow/laser status
(see below)
Status Value is encoded:
Flow
Laser
1
Good
Good
2
Good
Bad
3
Bad
Good
4
Bad
Bad
Note 1:
Option board 2 required
Note 2:
OPC Control setting enabled
Note 3:
(1) & (2) & External digital setting enabled
Table
E-1
Data Tags (Continued)
Tag Name
Type
Units
Access
Comment