Page D-8
IsoAir 310P Aerosol Particle Sensor Operations Manual
Appendix D Modbus
Associated Values for Specific Registry Entries
Associated Values for Specific Registry Entries
Table
D-5
Associated values for specific registry entries
Device Status
Entries
Device State
Entries (LSByte)
Device SubState
Entries (MSByte)
Data Packet Status
Entries
0x0001
Data Collection
Matches the common coil settings
0x0002
Data Available
0x0004
Data Clear
0x0008
Reset
0x0010
Green Status
0x0020
Red Status
0x0040
External Alarm
0x0080
IEEE Float mode
0
Idle
1
Sampling
2
Maintenance
3
Error
0
Idle
n/a
1
Idle
Start pump
2
Idle
Pre-Sample
0
Sampling
n/a
1
Sampling
Tare
0
Maintenance
n/a
1
Maintenance
Calibration
0
Error n/a
1
Error Leak
2
Error Cap
Off
0x0001
Laser Good
0x0002
Flow Good
0x0004
Hardware Good