Page 3-2
IsoAir 310P Aerosol Particle Sensor Operations Manual
Chapter 3 Configuring for Facility Net
Configuring the Sensor
Configuring the Sensor
CAUTION
As with any network-capable software, do not attempt to connect the IsoAir
310P particle sensor to a network without network administrator permissions.
>> To set configuration parameters:
1
Using the straight-through serial cable, connect the computer to the serial port on
the IsoAir 310P particle sensor.
NOTE:
Do NOT use a null modem cable.
2
Start the IsoAir 310P particle sensor.
3
Start the computer.
4
Start HyperTerminal.
5
Use HyperTerminal to set the following parameters:
Table
3-2
HyperTerminal Configuration
Baud
9600
Data Bits
8
Parity
N (none)
Stop Bits
1
Flow control
None
Parity Check
Off
Carrier Detect
Off
Connector
Appropriate to hardware (COM1, for example)