IsoAir 310P Aerosol Particle Sensor Operations Manual
Page 8-1
Chapter 8
Maintenance
Sterilization of Cleanroom
The IsoAir 310P particle sensor is enclosed in a sturdy, stainless steel enclosure and
should provide many years of trouble-free performance. Periodically it may be helpful
to wipe the case with a clean, lint-free cloth moistened with deionized water or a mild
cleanser.
The sample probe must be capped before the instrument is sterilized by
means of liquid, spray, or mist, ideally. Turn the pump
OFF
before
capping the sample probe.
Ingress of liquid or cleaning chemical may damage the internals of the sensor. Care
must be taken during cleaning processes.
Calibration, Service, and Shipping
The electronics module is highly sensitive to electrostatic discharge. This module has
a protective cover to prevent damage. When removing for calibration, perform the
following steps.
The instrument
MUST
be disconnected from its power source
BEFORE
opening the enclosure
>>
To remove the electronics module:
1
Follow the steps in Removing the Electronics Module on page
WARNING
WARNING