Sensor Installation
Chapter 2 Unpacking and Installation
IsoAir 310P Aerosol Particle Sensor Operations Manual
Page 2-9
Drilling Holes for Conduit Fittings (as needed)
The enclosure is pre-drilled and fitted with cable gland connectors. If a conduit
connection is needed, remove the cable gland connectors and use the pre-drilled
holes for connecting the conduit.
Exhaust
Cable
Glands
Figure 2-7
Conduit Port Placement, Dual
Mounting the Sensor
The IsoAir 310P particle sensor should be mounted to a wall or other vertical surface
that is capable of supporting a static load of 80 pounds (36.3 kg). It must be installed
by qualified personnel.
Because the electronics module is highly sensitive to electrostatic discharge, the
electronics module MUST be removed from its stainless steel enclosure before
attaching the stainless-steel enclosure to a wall or other vertical surface. See
Removing the Electronics Module on page
The enclosure can also be attached to a laboratory structure material, such as
Unistrut
®
, using similar attachment methods as for mounting to a metal-framed
hollow wall or solid-metal sheet wall.
CAUTION
The mounting location MUST be sufficiently strong. Do NOT mount the
instrument to un-reinforced sheetrock.