74
b) Call up the list of fixed tare values for taring:
In the Tare functions level, press the F1 Tare Fix key. The display shows the last fixed tare value called up.
Selection within the list
Scroll with the arrow keys.
Call the desired fixed value with F6 Accept.
Direct access via memory location number.
Enter the memory location number with the numeric keypad.
Direct access via barcode or EDP
(information in chapter 7.10 Barcode management or ask your Soehnle Industrial Solutions service partner)
Note
If the fixed memory recall via barcode function is activated and a fixed value not contained in the fixed
memory is recalled, the error message "Error 83 - Fixed value not available" is displayed. You are returned to
the weighing mode and the selected function is deleted.
Clear fixed tare value
To delete the tare value and the designation from the working memory, press the CL key.
6.5.12 Error messages when taring
Error 06:
Taring not possible.
Possible causes:
> Empty scale,
> Scale in overload or underload,
> Tare hand input with tare value over maximum load,
> Scale shows a negative weight value.
6.5.13 Display options with the Info key
Pressing the key gives you the following choice:
>
F1 Tare FixDisplay of
the stored tare fixed values (tare fix memory).
>
tare key
Display of the current tare weight in the tare memory.
After pressing the key , press the key .