7. Error messages
Error message
Cause /Aid / Info
4 REF Weight too small
Scale unsuitable for part weight, if necessary change the
setting in the setting mode
5 zeroing not possible
Deviation from zero point too large
6 Taring not possible
Empty scale, scale in under- or overload, tare hand input with
tare value over maximum load
7 Printing not permitted
Selected function does not allow to print, change or extend
function.
8 No unit switching
Shift key locked or disabled, no 2nd unit defined.
11 No measuring point signal
Cable interruption or short circuit to load cell, measuring point
defective
23 Printing failed
Printer not plugged in or off-line, cable connection missing or
defective
26 Negative value!
Wrong entry, delete entry and re-enter
27 Out of tolerance
Weight value is outside the set tolerance
45 totals Protocol memory full
Max. 999 entries possible, delete total memory
46 Cancellation not possible
Function blocked, wrong input in program sequence
47 Clear the totalizing memory first The totalizing memory must be cleared before a new
reference weight is determined; in setting mode the function
"Clear total before new reference weight" is activated.
48 No items available
Cancellation not possible, as there are no items in the
totalizing memory
49 Totalizing memory overflow
Value in totalizing memory too large, clear memory
50 Value below zero setting limit
Measured value lower than stored zero point, scale
mechanically damaged, apply preload, readjust, replace load
cell.
51 Value above zero setting limit
Measured value greater than stored zero point, scale
mechanically damaged, remove preload, readjust, replace
load cell.
52 Overload
Measured value above weighing range, unload the scale, load
cell defective
53 Underload
Scale with preload switched on and then preload removed,
switch off display on the keyboard, restore the weighing
platform to its original condition, switch on display again
75 Replace buffer battery
Replace battery
82 Number already available
The identifier assigned to a fixed value already exists
83 Fixed value not available
Correct input, check keyboard
90 Read-only memory missing / full A maximum of 999 memory locations are available per
memory
Your dealer and service partner or the service department of Soehnle Industrial Solutions will be happy to
assist you with further error analysis.