APPENDIX D
4618-021
120
Valid message identifiers
Table D-1 Table of identifiers
Command
1 2
Direction
3
Description
Type
4
‘A’ m
out
Bargraph reading
m = A signed number representing the value being sent to the
bargraph of the detector. A four digit number in the range
[±0..2047].
5
X
‘a’ n
‘a?’
in/out
in
Operator access
n = 0 : LIMITED
n = 1 : FULL
Operator access ?
X
‘B’ n
‘B?’
in/out
in
Buzzer enable
n = 0 : NO
n = 1 : YES
Buzzer enable ?
X
‘b’
-----------
SEE LATER FOR DETAILS OF IDENTIFIER ‘b’
X
‘C’ n ‘,’ s1 ‘,’ c1
‘,’ t ‘,’ a
‘C?’ n
in/out
in
Preset Product Settings.
n = Product number, two digit number, in the range [0..20]
s1 = Sensitivity, three digit number, in the range [0..199]
c1 = Phase setting, five digit fixed point number, in the range
[000.00 ... 180.00], resolution is 0.02.
t = Timer group
t = 0 : Timer group A
t = 1 : Timer group B
t = 2 : Timer group C
a = Frequency offset , single digit number in the range [0..1].
Note. If parameter n is 0 then c1 must be 00.00
Product Settings ?
P
'c'n
'c?'
in/out
Printer Handshake
n = 0 : HW
n = 1 : SW
Printer Handshake?
X
‘D’ d ‘,’ c
‘D’ d ‘,’ c
‘D?’
out
in
in
Counter values.
d = reject counter, four digit number, in the range [0..9999]
c = pack counter, eight digit number, in the range [0..16777214]
d = 1 : Reset reject counter
c = 1 : Reset pack counter
Counter Values ?
X
‘d’ s ‘,’ m
‘d?’
out
in
Serial number and Model number of the detector.
s = Head serial number of the device, five digit number, in the
range [0..65535].
s is read only, any value transmitted to the detector is ignored.
m = Model number of the device, a six digit number.
Note: Only the last five digits can be set by the host. The first
digit is ignored by the detector when received.
(See instruction manual for details of this number)
Serial and model number ?
F
‘E’
---------
SEE LATER FOR DETAILS OF IDENTIFIER ‘E’
X
Summary of Contents for Signature
Page 16: ......
Page 26: ...INSTALLATION 10 4618 021 Fig 5 Typical Metal Detector Conveyor ...
Page 31: ...INSTALLATION 4618 021 15 Fig 10 Cable Gland Assemblies for Power Supply Unit ...
Page 52: ...OPERATING INSTRUCTIONS 36 4618 021 Fig 14 Supervisor Mode Flow Chart ...
Page 55: ...OPERATING INSTRUCTIONS 4618 021 39 Fig 15 Engineer Mode Flow Chart ...
Page 71: ...OPERATING INSTRUCTIONS 4618 021 55 Fig 16 Pack Sensor Mode Flow Chart ...
Page 82: ...OPERATING INSTRUCTIONS 66 4618 021 Fig 18 QA Inspector And Operator Flow Chart ...
Page 92: ...OPERATING INSTRUCTIONS 76 4618 021 Fig 19 Viewing Mode Flow Chart ...