OPERATING INSTRUCTIONS
92
4618-021
18) Press the soft key adjacent to REJECT TIME.
a) Place the metal contaminant at the front or
rear edge of the product package.
b) Pass the product with the metal contaminant
through the detector aperture, ensuring that
the product does not slip or skew.
c) Use the cursor Up and Down keys to adjust
the Reject Time until the pack is rejected
accurately and repeatedly.
19) Press the Enter key to accept the new settings.
20) Press the Enter key a second time to proceed to
the next display.
21) The display shown opposite will appear for five
seconds before returning to the Engineer Mode
display.
Testing the Reject Device.
22)
a) Check the setting of tm2G by passing the
product and metal contaminant through the
aperture of the detector.
b) Check the reject device operates correctly
with the metal contaminant at both the front
and rear of the pack.
c) Check the reject device operates correctly
with different sizes of metal contaminant.
Under normal operating conditions with packs
continuously passing along a conveyor if the
gap between a metal contaminated pack and
the adjacent pack is less than half the pack
length both packs may be rejected.
23) If for any reason the calibration of the tm2G
timer is not correct return to step 1) and repeat
all the steps.
Summary of Contents for Signature
Page 16: ......
Page 26: ...INSTALLATION 10 4618 021 Fig 5 Typical Metal Detector Conveyor ...
Page 31: ...INSTALLATION 4618 021 15 Fig 10 Cable Gland Assemblies for Power Supply Unit ...
Page 52: ...OPERATING INSTRUCTIONS 36 4618 021 Fig 14 Supervisor Mode Flow Chart ...
Page 55: ...OPERATING INSTRUCTIONS 4618 021 39 Fig 15 Engineer Mode Flow Chart ...
Page 71: ...OPERATING INSTRUCTIONS 4618 021 55 Fig 16 Pack Sensor Mode Flow Chart ...
Page 82: ...OPERATING INSTRUCTIONS 66 4618 021 Fig 18 QA Inspector And Operator Flow Chart ...
Page 92: ...OPERATING INSTRUCTIONS 76 4618 021 Fig 19 Viewing Mode Flow Chart ...