OPERATING INSTRUCTIONS
4618 - 021
91
10) Ensure that the pack sensor beam is clear of
any packs or products.
11) Select the soft key adjacent to WINDOW TIME.
The display will show ‘-RUN-’.
12) Pass the product sample through the pack
sensor beam, ensuring that it does not skew or
slip on the conveyor.
The metal detector will automatically measure the
window time and display this instead of the ‘-
RUN
-’
message.
13) Repeat steps 11) and 12) to ensure that the
window time is being measured correctly and is
repeatedly calculating the same value, (within
±
10%).
14) Ensure that the pack sensor beam is clear of
any packs or products.
15) Select the soft key adjacent to the SYNC.
DELAY.
When the Sync. delay is selected the detect led and
buzzer will sound permanently
.
a) Pass the product through the detector
aperture, ensuring that the product does not
slip or skew.
b) Use the cursor Up and Down keys to adjust
the Sync Delay so that the pack is rejected
accurately and consistently.
16) Press the Enter key to accept the new settings.
17) Press the soft key adjacent to ETC.
Summary of Contents for Signature
Page 16: ......
Page 26: ...INSTALLATION 10 4618 021 Fig 5 Typical Metal Detector Conveyor ...
Page 31: ...INSTALLATION 4618 021 15 Fig 10 Cable Gland Assemblies for Power Supply Unit ...
Page 52: ...OPERATING INSTRUCTIONS 36 4618 021 Fig 14 Supervisor Mode Flow Chart ...
Page 55: ...OPERATING INSTRUCTIONS 4618 021 39 Fig 15 Engineer Mode Flow Chart ...
Page 71: ...OPERATING INSTRUCTIONS 4618 021 55 Fig 16 Pack Sensor Mode Flow Chart ...
Page 82: ...OPERATING INSTRUCTIONS 66 4618 021 Fig 18 QA Inspector And Operator Flow Chart ...
Page 92: ...OPERATING INSTRUCTIONS 76 4618 021 Fig 19 Viewing Mode Flow Chart ...