OPERATING INSTRUCTIONS
94
4618-021
tm3[G] TIMER - NON GATED MODE
!
CAUTION
Ensure you have read the REJECT TIMERS
INTRODUCTION section on page 78
Introduction
The tm3 timer is used on variable speed product lines
or product lines that may be stopped or started by
external equipment.
The tm3 timer when used in the non-gated mode is
ideal for use with loose product where the reject
device is situated some distance from the detector.
This timer is also suitable for use with packaged
products but is not as accurate as the gated version,
tm3G.
NOTE
Remember to record all settings on the Settings
Sheets at the rear of the manual.
This timer type is suitable for reject devices which
include plough-action diverters, retracting bands etc.
Setting up Timer Type tm3
The display opposite may be used to either:
1)
Select which timer group the current product
number should use,
or,
2)
Select which timer group the current product
number should use AND adjust the timer group
settings such as timer type, reject shift, signal
shift etc.
NOTE
REMINDER
If you have already selected the timer group and
pressed YES to updating the group, as per the
Engineer Mode section earlier, then proceed to
step 3).
1)
Press the soft key adjacent to the required group
letter, A, B or C.
NOTE
EXPLANATORY NOTE
The tm3 and tm3G timer modes continually refer to
terms such as 'SIGNAL SHIFT', 'REJECT SHIFT',
'SYNC SHIFT' etc. The word shift when used in these
context refers to the distance the product will move (or
shift) between successive pulses on the speed sensor
input. Thus if a speed sensor has a shift of 10 cm
then a REJECT SHIFT of 5 is equivalent to a product
moving a distance of 5x10 cm = 50 cm.
For instructions on calculating the shift distance for
your application refer to Appendix A.
tm3 settings
1)
Signal Shift
This is the number of speed sensor pulses taken
for a metal contaminant to travel from the
detector to the reject device.
2)
Reject Shift
This is the number of speed sensor pulses
required by the reject device to reject the
contaminated product.
The currently selected timer group will be indicated by a flashing
cursor appearing over the group letter.
Summary of Contents for Signature
Page 16: ......
Page 26: ...INSTALLATION 10 4618 021 Fig 5 Typical Metal Detector Conveyor ...
Page 31: ...INSTALLATION 4618 021 15 Fig 10 Cable Gland Assemblies for Power Supply Unit ...
Page 52: ...OPERATING INSTRUCTIONS 36 4618 021 Fig 14 Supervisor Mode Flow Chart ...
Page 55: ...OPERATING INSTRUCTIONS 4618 021 39 Fig 15 Engineer Mode Flow Chart ...
Page 71: ...OPERATING INSTRUCTIONS 4618 021 55 Fig 16 Pack Sensor Mode Flow Chart ...
Page 82: ...OPERATING INSTRUCTIONS 66 4618 021 Fig 18 QA Inspector And Operator Flow Chart ...
Page 92: ...OPERATING INSTRUCTIONS 76 4618 021 Fig 19 Viewing Mode Flow Chart ...