Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
130
User Manual 1178.5989.02 ─ 06
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-10 dBm
Default unit: DBM
Example:
TRIG:LEV:IFP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IQPower
<TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event.
Note that any RF attenuation is considered when the trigger level is analyzed. If
defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
Range:
-130 dBm to 30 dBm
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IQP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:SLOPe
<Type>
Parameters:
<Type>
POSitive | NEGative
POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST:
POSitive
Example:
TRIG:SLOP NEG
Manual operation:
See
TRIGger[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
For details on trigger sources see
Note on external triggers:
Configuring I/Q Analyzer Measurements