Sweep Points
In the I/Q Analyzer application, a specific frequency bandwidth is swept for a specified
measurement time. During this time, a defined number of samples (= "Record
Length" ) are captured. These samples are then evaluated by the applications. There-
fore, in this case the number of sweep points does not define the amount of data to be
acquired, but rather the number of trace points that are evaluated and displayed in the
result diagrams.
Note:
For some result displays, the sweep points may not be editable as they are
determined automatically, or restrictions may apply. For the I/Q vector result display,
the number of I/Q samples to record ( "Record Length" ) must be identical to the num-
ber of trace points to be displayed ("Sweep Points"). Thus, the sweep points are not
editable for this result display. If the "Record Length" is edited, the sweep points are
adapted automatically. For record lengths outside the valid range of sweep points, i.e.
less than 101 points or more than 100001 points, the diagram does not show valid
results.
Using fewer than 4096 sweep points with a detector other than "Auto Peak" may lead
to wrong level results. For details see
"Combining results - trace detector"
Remote command:
[SENSe:]SWEep[:WINDow<n>]:POINts
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average" , "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 sweeps. For "Sweep Count" =1, no averaging, maxhold or minhold operations
are performed.
Data Acquisition and Bandwidth Settings