Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
154
User Manual 1178.5989.02 ─ 06
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
[SENSe:]SWEep:COUNt:CURRent?
This query returns the current number of started sweeps or measurements. This com-
mand is only available if a sweep count value is defined and the instrument is in single
sweep mode.
Return values:
<CurrentCount>
Example:
SWE:COUNt 64
Sets sweep count to 64
INIT:CONT OFF
Switches to single sweep mode
INIT
Starts a sweep (without waiting for the sweep end!)
SWE:COUN:CURR?
Queries the number of started sweeps
Usage:
Query only
[SENSe:]SWEep[:WINDow<n>]:POINts
<SweepPoints>
This command defines the number of sweep points to analyze after a sweep.
Capturing Data and Performing Sweeps