![Rohde & Schwarz ZNL Series User Manual Download Page 109](http://html.mh-extra.com/html/rohde-and-schwarz/znl-series/znl-series_user-manual_1477893109.webp)
Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
109
User Manual 1178.5989.02 ─ 06
●
All measurements from the previous application (e.g. Spectrum) are turned off
●
All traces are set to "Blank" mode
●
The I/Q data analysis mode is turned off (
TRAC:IQ:EVAL OFF
, if previous applica-
tion was also I/Q Analyzer)
Note:
To turn trace display back on or to enable the evaluation functions of the I/Q
Analyzer, execute the
TRAC:IQ:EVAL ON
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
TRAC:IQ ON
Switches on I/Q data acquisition
9.4
Configuring I/Q Analyzer Measurements
The following commands configure the I/Q Analyzer measurements.
Configuring the Data Input and Output
................................................................. 109
Configuring the Vertical Axis (Amplitude, Scaling)
................................................121
.................................................................................134
Adjusting Settings Automatically
...........................................................................140
9.4.1
Configuring the Data Input and Output
The following commands are required to configure data input and output.
....................................................................................... 111
.................................................................................111
........................................................................................ 119
9.4.1.1
RF Input
INPut<ip>:ATTenuation:PROTection:RESet
.......................................................................110
Configuring I/Q Analyzer Measurements