![Rohde & Schwarz ZNL Series User Manual Download Page 89](http://html.mh-extra.com/html/rohde-and-schwarz/znl-series/znl-series_user-manual_1477893089.webp)
Analysis
R&S
®
ZNL
89
User Manual 1178.5989.02 ─ 06
All markers can be defined as band power markers, each with a different span. When a
band power marker is activated, if no marker is active yet, marker 1 is activated. Other-
wise, the currently active marker is used as a band power marker (all other marker
functions for this marker are deactivated).
If the detector mode for the marker trace is set to "Auto", the RMS detector is used.
The individual marker settings correspond to those defined in the "Marker" dialog box
(see
Chapter 6.4.1.1, "Individual Marker Setup"
on page 72). Any settings to the
marker state or type changed in the "Marker Function" dialog box are also changed in
the "Marker" dialog box and vice versa.
Marker Usage